ProSpore 2
ProSpore2 is a self contained Biological Indicator for use in
either Steam or Ethylene Oxide (EO) sterilization. Each unit is
made up of a plastic container with cap, a crushable glass ampoule
with recovery media, and a paper disc inoculated with spores.
ProSpore2 EO
Suitable for use in hospital or industrial Ethylene Oxide
sterilization.
Contains B. atrophaeus spores
Catalog # PS2-1-6-50 (106)
Intended for use in EO 600mg/L, 54°C, 60% RH
Sterilization failure evident by a dramatic green to yellow
color change and/or turbidity.
|
 |
ProSpore2 Steam
Suitable for use in hospital or industrial Steam
sterilization.
Contains G.
stearothermophilus1 spores
Catalog # PS2-3-5-50 (105)
PS2-3-6-50 (106)
Sterilization failure evident by a dramatic purple to
yellow color change and/or turbidity.
|
Do not depress cap prior to exposure in the sterilizer. Allow unit
to cool after exposure, depress cap completely and crush ampoule.
Incubation
ProSpore2 for Steam has an incubation period of 24 hours. ProSpore2
for EO has an incubation period of 48-hours. ProSpore2 for Steam
should be incubated at 55-66°C. ProSpore2 for EO should be
incubated at 30-35°C.
Labeling
Each unit is labeled with Lot #, Expiration, Manufacturer and type
of sterilization in which it should be used. Space is allotted on
the label for the user to write the date, sterilizer #, and load
information. Label has a chemical indicator bar printed on it to
assist the user in distinguishing an exposed from an unexposed
unit. Box label contains similar information, including species,
storage conditions, and manufacturer contact information.
ProSpore2 is packaged in 50 units per box.
Certification
Each box of ProSpore2 is certified for Population, Species,
D-value, z-value (where appropriate) and expiration date. Shelf
life is 18 months from the date of manufacture.
 |